Vulnerabilities (CVE)

Filtered by vendor Baxter Subscribe
Total 27 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2024-6795 1 Baxter 1 Connex Health Portal 2024-09-20 N/A 9.8 CRITICAL
In Connex health portal released before8/30/2024, SQL injection vulnerabilities were found that could have allowed an unauthenticated attacker to gain unauthorized access to Connex portal's database.  An attacker could have submitted a crafted payload to Connex portal that could have resulted in modification and disclosure of database content and/or perform administrative operations including shutting down the database.
CVE-2024-6796 1 Baxter 1 Connex Health Portal 2024-09-20 N/A 9.1 CRITICAL
In Baxter Connex health portal released before 8/30/2024, an improper access control vulnerability has been found that could allow an unauthenticated attacker to gain unauthorized access to Connex portal's database and/or modify content.
CVE-2022-26394 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-02-04 N/A 5.4 MEDIUM
The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host. This may allow an attacker to perform a man in the middle attack that modifies parameters making the network connection fail.
CVE-2022-26392 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-02-04 N/A 6.5 MEDIUM
The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information.
CVE-2022-26393 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-02-04 N/A 8.1 HIGH
The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information or cause a Denial of Service (DoS) on the WBM.
CVE-2022-26390 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-02-04 N/A 4.2 MEDIUM
The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased may be able to extract sensitive information.
CVE-2021-43935 1 Baxter 10 Welch Allyn Connex Cardio, Welch Allyn Diagnostic Cardiology Suite, Welch Allyn Hscribe Holter Analysis System and 7 more 2024-02-04 6.8 MEDIUM 9.8 CRITICAL
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.
CVE-2020-12048 1 Baxter 2 Phoenix X36, Phoenix X36 Firmware 2024-02-04 5.0 MEDIUM 7.5 HIGH
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
CVE-2020-12047 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-02-04 7.5 HIGH 9.8 CRITICAL
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.
CVE-2020-12020 1 Baxter 4 Em1200, Em1200 Firmware, Em2400 and 1 more 2024-02-04 3.6 LOW 6.1 MEDIUM
Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an attacker to alter the startup script as the limited-access user.
CVE-2020-12036 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2024-02-04 5.0 MEDIUM 7.5 HIGH
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVE-2020-12037 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2024-02-04 5.0 MEDIUM 7.5 HIGH
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVE-2020-12043 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-02-04 7.5 HIGH 9.8 CRITICAL
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted.
CVE-2020-12045 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-02-04 7.5 HIGH 9.8 CRITICAL
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.
CVE-2020-12035 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2024-02-04 3.6 LOW 4.9 MEDIUM
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.
CVE-2020-12016 1 Baxter 4 Em1200, Em1200 Firmware, Em2400 and 1 more 2024-02-04 10.0 HIGH 9.8 CRITICAL
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI.
CVE-2020-12039 1 Baxter 2 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware 2024-02-04 2.1 LOW 2.4 LOW
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
CVE-2020-12012 1 Baxter 4 Em1200, Em1200 Firmware, Em2400 and 1 more 2024-02-04 3.6 LOW 6.1 MEDIUM
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
CVE-2020-12032 1 Baxter 4 Em1200, Em1200 Firmware, Em2400 and 1 more 2024-02-04 6.4 MEDIUM 9.1 CRITICAL
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including PHI.
CVE-2020-12008 1 Baxter 4 Em1200, Em1200 Firmware, Em2400 and 1 more 2024-02-04 5.0 MEDIUM 7.5 HIGH
Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including PHI.