Vulnerabilities (CVE)

Filtered by vendor Medtronic Subscribe
Filtered by product Minimed Paradigm 512
Total 1 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2019-10964 1 Medtronic 38 Minimed 508, Minimed 508 Firmware, Minimed Paradigm 511 and 35 more 2025-05-22 5.8 MEDIUM 7.1 HIGH
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.