Vulnerabilities (CVE)

Filtered by vendor Medtronic Subscribe
Filtered by product Carelink Monitor 2490c Firmware
Total 1 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2019-6540 1 Medtronic 46 Amplia Crt-d, Amplia Crt-d Firmware, Carelink 2090 and 43 more 2024-11-21 3.3 LOW 6.5 MEDIUM
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.