Total
2 CVE
CVE | Vendors | Products | Updated | CVSS v2 | CVSS v3 |
---|---|---|---|---|---|
CVE-2019-10962 | 1 Bd | 2 Alaris Gateway Workstation, Alaris Gateway Workstation Firmware | 2024-11-21 | 5.0 MEDIUM | 5.3 MEDIUM |
BD Alaris Gateway versions, 1.0.13,1.1.3 Build 10,1.1.3 MR Build 11,1.1.5, and 1.1.6, The web browser user interface on the Alaris Gateway Workstation does not prevent an attacker with knowledge of the IP address of the Alaris Gateway Workstation terminal to gain access to the status and configuration information of the device. | |||||
CVE-2019-10959 | 1 Bd | 10 Alaris Cc Syringe Pump, Alaris Cc Syringe Pump Firmware, Alaris Gateway Workstation and 7 more | 2024-11-21 | 7.5 HIGH | 10.0 CRITICAL |
BD Alaris Gateway Workstation Versions, 1.1.3 Build 10, 1.1.3 MR Build 11, 1.2 Build 15, 1.3.0 Build 14, 1.3.1 Build 13, This does not impact the latest firmware Versions 1.3.2 and 1.6.1, Additionally, the following products using software Version 2.3.6 and below, Alaris GS, Alaris GH, Alaris CC, Alaris TIVA, The application does not restrict the upload of malicious files during a firmware update. |