CVE-2022-36385

A threat actor with momentary access to the device can plug in a USB drive and perform a malicious firmware update, resulting in permanent changes to device functionality. No authentication or controls are in place to prevent a threat actor from maliciously modifying firmware and performing a drive-by attack to load the firmware on any CMS8000 device.
References
Link Resource
https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 Mitigation Third Party Advisory US Government Resource
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*
cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*

History

15 Sep 2022, 17:27

Type Values Removed Values Added
CWE NVD-CWE-noinfo
CPE cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*
cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*
CVSS v2 : unknown
v3 : unknown
v2 : unknown
v3 : 6.8
References (MISC) https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 - (MISC) https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01 - Mitigation, Third Party Advisory, US Government Resource

13 Sep 2022, 15:45

Type Values Removed Values Added
New CVE

Information

Published : 2022-09-13 15:15

Updated : 2024-02-04 22:51


NVD link : CVE-2022-36385

Mitre link : CVE-2022-36385

CVE.ORG link : CVE-2022-36385


JSON object : View

Products Affected

contechealth

  • cms8000
  • cms8000_firmware
CWE
NVD-CWE-noinfo CWE-284

Improper Access Control