CVE-2015-3952

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
References
Link Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 Mitigation US Government Resource Third Party Advisory
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:pifzer:plum_a\+_infusion_system_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:pifzer:plum_a\+_infusion_system:-:*:*:*:*:*:*:*

Configuration 2 (hide)

AND
cpe:2.3:o:pifzer:plum_a\+3_infusion_system_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:pifzer:plum_a\+3_infusion_system:-:*:*:*:*:*:*:*

Configuration 3 (hide)

AND
cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*

History

No history.

Information

Published : 2019-03-25 16:29

Updated : 2024-02-04 20:20


NVD link : CVE-2015-3952

Mitre link : CVE-2015-3952

CVE.ORG link : CVE-2015-3952


JSON object : View

Products Affected

pifzer

  • plum_a\+3_infusion_system_firmware
  • plum_a\+_infusion_system
  • plum_a\+_infusion_system_firmware
  • symbiq_infusion_system
  • symbiq_infusion_system_firmware
  • plum_a\+3_infusion_system
CWE
CWE-200

Exposure of Sensitive Information to an Unauthorized Actor

CWE-312

Cleartext Storage of Sensitive Information