CVE-2015-1012

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
References
Link Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 Mitigation US Government Resource Third Party Advisory
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:pfizer:lifecare_pca_infusion_system_firmware:*:*:*:*:*:*:*:*
cpe:2.3:h:pfizer:lifecare_pca_infusion_system:-:*:*:*:*:*:*:*

History

No history.

Information

Published : 2019-03-25 19:29

Updated : 2024-02-04 20:20


NVD link : CVE-2015-1012

Mitre link : CVE-2015-1012

CVE.ORG link : CVE-2015-1012


JSON object : View

Products Affected

pfizer

  • lifecare_pca_infusion_system_firmware
  • lifecare_pca_infusion_system
CWE
CWE-200

Exposure of Sensitive Information to an Unauthorized Actor

CWE-312

Cleartext Storage of Sensitive Information