An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
References
Link | Resource |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | Mitigation Third Party Advisory US Government Resource |
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 | Mitigation Third Party Advisory US Government Resource |
Configurations
Configuration 1 (hide)
AND |
|
History
21 Nov 2024, 02:12
Type | Values Removed | Values Added |
---|---|---|
References | () https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 - Mitigation, Third Party Advisory, US Government Resource |
Information
Published : 2019-03-26 16:29
Updated : 2024-11-21 02:12
NVD link : CVE-2014-5433
Mitre link : CVE-2014-5433
CVE.ORG link : CVE-2014-5433
JSON object : View
Products Affected
baxter
- wireless_battery_module
- sigma_spectrum_infusion_system
- sigma_spectrum_infusion_system_firmware